Akan Industrial and Automation

Multisensor Systems

Multisensor MS4000-S VOC Gas Analyzer

Online VOC and hydrocarbon gas analyzer (1–3000 ppm) for drains, effluent and sewer headspace, gas streams, fuel leaks and odour control, with a reading every 2 seconds.

  • Drains and effluent discharge headspace
  • Sewer system monitoring
  • VOCs in gas streams and gas removal systems
  • Oil and fuel leak detection
  • Odour control
Multisensor MS4000-S VOC Gas Analyzer

The Multisensor MS4000-S is an online VOC and hydrocarbon gas analyzer that measures volatile organic compounds in gas from 1 to 3000 ppm, with a new reading every 2 seconds. It is used on drains, effluent discharges and sewer systems, on gas streams and gas removal systems, and for fuel leak detection and odour control. Akan Industrial and Automation supplies and commissions the MS4000-S in Türkiye as Multisensor Systems dealer.

How does it work?

The MS4000-S continuously draws air from above the effluent, from inside a drain or from a gas line and passes it across its sensor, which responds to VOCs. The concentration is shown on the touchscreen and sent to a PLC or SCADA system over 4–20 mA. A user-programmable alert threshold switches relays and front-panel indicators, so the analyzer can directly trigger isolation valves, pumps, samplers or audible and visual alarms.

Typical applications

Industrial effluent

Spills can overload on-site treatment and lead to illegal discharges. An MS4000-S at the outlet gives an immediate warning when VOC levels rise, so that action can be taken before the discharge leaves the site.

Sewer systems

Hydrocarbons in sewers endanger maintenance staff and can contaminate sewage sludge, which must then be disposed of as chemical waste instead of being reused. Detecting VOC peaks in the network helps find the source and protect the treatment plant.

Gas streams and leak detection

The analyzer also monitors VOCs in gas removal systems and exhaust gas, detects leaks in natural gas extraction wells and is used for odour control and land remediation.

Key specifications

Parameter Value
Detection range 1–3000 ppm total VOCs (default 4–20 mA range 0–3000 ppm)
Repeatability ±5%
Accuracy ±15% (100 ppm toluene standard)
Measurement Continuous, new reading every 2 seconds
Ambient temperature 0–50 °C
Sample humidity Up to 90% RH, non-condensing
Outputs 4–20 mA (isolated), 3 relays: alarm 1, alarm 2, fault (5 A)
Communication Modbus optional; Profibus, HART, Foundation Fieldbus on request
Interface and data Touchscreen; data storage on µSD card / PC (6 months)
Enclosure IP65 powder-coated stainless steel; power supply module IP68
Dimensions and weight 45 × 28 × 20 cm, 15 kg
Power supply 90–240 V AC 50/60 Hz or 9–36 V DC; max 45 W, typically 20 W
Maintenance Service every 6 months (filters, restrictors)

To measure VOCs dissolved in water rather than in gas, see the MS1200-S, MS1210-S and MS1220-S. For hazardous-area installations, discuss the location with us before selecting the analyzer.

Why choose Akan for the MS4000-S?

Akan plans the sampling point and sample line, integrates the relays and 4–20 mA signal into your control system and commissions the analyzer on site. Contact us with your application and target compounds.

Frequently asked questions

What does the MS4000-S measure?

It measures the total concentration of volatile organic compounds in a gas sample, from 1 to 3000 ppm depending on the target gas, reported against a toluene standard.

How is the MS4000-S used on drains and effluent?

It continuously samples the air above the effluent or inside the drain and passes it across the sensor. A rise in VOCs indicates a spill, so the relay can stop pumps, close valves or start a sampler.

Which gases can it be set up for?

Common targets are petroleum gas, methane, isobutane, kerosene, ethanol, diesel, isobutylene, toluene, acetone, xylene, heptane, benzene, methanol and mercaptans.

What is the difference between the MS4000-S and the MS1210-S?

The MS1210-S measures VOCs dissolved in water through a sampling tank and is more accurate; the MS4000-S samples gas directly above a drain or from a gas stream, which is a lower-cost solution with lower absolute accuracy.

Request a quote

Tell us about your application and our engineers will get back to you within one business day.